Ensuring the safety of a centrifuge is the top priority, as losing control of the centrifugal force can lead to serious damage. It is crucial to check that the centrifuge tubes are properly balanced, that the rotational speed does not exceed the set limit, and that the rotor is in good condition without any signs of corrosion or wear.
Equipment Required: High-speed refrigerated centrifuge, rotor, and centrifuge tubes.
Separation Technology Operation Steps:
- Balance the centrifuge tubes carefully before use.
- Follow the correct procedure for operating the centrifuge.
- Learn proper maintenance techniques for both the centrifuge and its rotor.
Key Considerations When Using a Centrifuge:
- Always complete the registration form before using the centrifuge. Record details such as the user’s name, type of rotor, speed, and time.
- Ensure the centrifuge tubes are perfectly balanced. Pay close attention to the positioning of the rotor to maintain balance. Never exceed the maximum speed rating of the centrifuge or rotor.
- Do not leave the centrifuge unattended once it has reached the preset speed. If anything seems unusual, stop the machine immediately.
- You can often tell if the centrifuge is operating normally by listening to the sound and observing the level of vibration.
- After using high-salt solutions like ammonium sulfate, make sure to thoroughly clean the rotor and the centrifuge chamber to prevent residue buildup.
- Ultra-high-speed centrifuges operate at extremely high speeds and require specialized training before they can be used safely.
Step-by-Step Usage Guide:
It's important to develop the habit of registering your usage each time you use the centrifuge.
- Set the desired operating temperature (usually 4°C). Place the rotor into the centrifuge chamber, making sure it is securely clamped onto the shaft. Close the door and allow the chamber to cool down sufficiently before proceeding.
- Prepare your sample by placing it in a suitable centrifuge tube. Balance the weight using a balance scale. Cover and tighten the tube securely.
- Note: Fill the tube only up to 70% capacity. Even with a lid, excessive centrifugal force may cause leakage.
- Note: Most tubes have lids, so ensure that the lid is also balanced when placing them in the rotor.
- Note: Tubes are often placed on crushed ice. When balancing, wipe away any excess moisture from the ice.
- Note: Always balance the sample tube with another one filled with clean water to ensure symmetry.
- Place the well-balanced tubes symmetrically into the rotor and secure the rotor cover. Double-check that the cover is tightly fastened.
- Warning: If the rotor cover is not properly tightened, it may come loose during operation, causing severe damage.
- Close the centrifuge door and set the desired speed and time on the control panel.
By following these steps and maintaining a strong focus on safety, you can ensure efficient and safe operation of the centrifuge every time. Always stay alert and follow all manufacturer guidelines to avoid accidents and prolong the life of your equipment.